Title: | An ultra-sensitive ∆R/R measurement system for biochemical sensors using piezoresistive micro-cantilevers |
Authors: | Rao, V. Ramgopal |
Keywords: | EEE Biosensors Piezoresistance Electrical resistance measurement Operational amplifiers Immune system Signal generators Voltage |
Issue Date: | 2009 |
Publisher: | IEEE |
Abstract: | Piezoresistive micro-cantilevers are interesting bio-sensing tool whose base resistance value (R) changes by a few parts per million (DeltaR) in deflected conditions. Measuring such a small deviation is always being a challenge due to noise. An advanced and reliable DeltaR/R measurement scheme is presented in this paper which can sense resistance changes down to 6 parts per million. The measurement scheme includes the half-bridge connected micro-cantilevers with mismatch compensation, precision op-amp based filters and amplifiers, and a lock-in amplifier based detector. The input actuating sine wave is applied from a function generator and the output dc voltage is displayed on a digital multimeter. The calibration is performed and instrument sensitivity is calculated. An experimental set-up using a probe station is discussed that demonstrates a combined performance of the measurement system and SU8-polysilicon cantilevers. The deflection sensitivity of such polymeric cantilevers is calculated. The system will be highly useful to detect bio-markers such as myoglobin and troponin that are released in blood during or after heart attacks. |
URI: | https://ieeexplore.ieee.org/document/5334434 http://dspace.bits-pilani.ac.in:8080/xmlui/handle/123456789/12837 |
Appears in Collections: | Department of Electrical and Electronics Engineering |
Files in This Item:
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.