Please use this identifier to cite or link to this item:
http://dspace.bits-pilani.ac.in:8080/jspui/handle/123456789/20073| Title: | pH calibration of tetroxalate, tartrate and phthalate buffer solutions at above 100°C |
| Authors: | Manning, Gary D. |
| Keywords: | Chemistry Buffer solutions Tetroxalate buffer Journal of the Chemical Society : Faraday Transaction - I High-temperature pH calibration |
| Issue Date: | 1978 |
| Publisher: | Journal of the Chemical Society : Faraday Transaction - I. The Chemical Society, London. 1978, 74 (09-12) |
| Abstract: | The pHS of three standard buffer solutions has been measured using a hydrogen electrode against a silver–silver chloride electrode in a cell nominally without a liquid–liquid junction. The measurements have been carried out at 95–120°C in 0.05 mol kg–1 potassium tetroxalate, 100–160°C in saturated (25°C) potassium hydrogen tartrate and 100–200°C in 0.05 mol kg–1 potassium hydrogen phthalate solution. It has been shown that the method cannot be used to calibrate phosphate or borate buffers at >100°C because of excessive reduction of silver chloride by hydrogen, although such solutions remain useful buffers at elevated temperatures. |
| URI: | http://dspace.bits-pilani.ac.in:8080/jspui/handle/123456789/20073 |
| Appears in Collections: | Journal Articles (before-1995) |