dc.contributor.author |
Rao, V. Ramgopal |
|
dc.date.accessioned |
2023-11-03T07:02:34Z |
|
dc.date.available |
2023-11-03T07:02:34Z |
|
dc.date.issued |
2009 |
|
dc.identifier.uri |
https://ieeexplore.ieee.org/document/5334434 |
|
dc.identifier.uri |
http://dspace.bits-pilani.ac.in:8080/xmlui/handle/123456789/12837 |
|
dc.description.abstract |
Piezoresistive micro-cantilevers are interesting bio-sensing tool whose base resistance value (R) changes by a few parts per million (DeltaR) in deflected conditions. Measuring such a small deviation is always being a challenge due to noise. An advanced and reliable DeltaR/R measurement scheme is presented in this paper which can sense resistance changes down to 6 parts per million. The measurement scheme includes the half-bridge connected micro-cantilevers with mismatch compensation, precision op-amp based filters and amplifiers, and a lock-in amplifier based detector. The input actuating sine wave is applied from a function generator and the output dc voltage is displayed on a digital multimeter. The calibration is performed and instrument sensitivity is calculated. An experimental set-up using a probe station is discussed that demonstrates a combined performance of the measurement system and SU8-polysilicon cantilevers. The deflection sensitivity of such polymeric cantilevers is calculated. The system will be highly useful to detect bio-markers such as myoglobin and troponin that are released in blood during or after heart attacks. |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
IEEE |
en_US |
dc.subject |
EEE |
en_US |
dc.subject |
Biosensors |
en_US |
dc.subject |
Piezoresistance |
en_US |
dc.subject |
Electrical resistance measurement |
en_US |
dc.subject |
Operational amplifiers |
en_US |
dc.subject |
Immune system |
en_US |
dc.subject |
Signal generators |
en_US |
dc.subject |
Voltage |
en_US |
dc.title |
An ultra-sensitive ∆R/R measurement system for biochemical sensors using piezoresistive micro-cantilevers |
en_US |
dc.type |
Article |
en_US |